

43Impedance imbalance between the paired electrodes and movement of the electrodes can significantly distort or eliminate the electrocardiographic signal. 44Electrode-related specific problems include excessive offset potential and polarization (buildup of electrical charge at its base plate as a result of current flow). There may be an electrical potential across a pair of these electrodes equal to the algebraic difference of the two half-cell potentials, called the offset potential . 43Disposable electrodes develop ionic potentials at the electrode–electrolyte interface known as half-cell potentials. 42The surface electrodes convert an ionic current into a flow of electrons. 41Voltage of several millivolts can be generated by physically stretching the epidermis.

Ultimately, an improved understanding of the artifacts generated by equipment, and their identifying characteristics, is important to avoid misinterpretation, misdiagnosis, and iatrogenic complication.īecause of varying tissue resistances from heart to skin, attenuated body surface potentials have an amplitude of only 1% of the amplitude of transmembrane potentials (0.5 to 2.0 mV) across cardiac cells. Technological advancements in processing the electrocardiographic signal may be useful to detect and eliminate artifacts. Artefact is a not uncommon finding on electrocardiograms taken in outpatients, emergency departments, or intensive care settings. Other devices: If a laptop or tablet is not being used, or the AC adaptor was not the cause then the artifact is being caused by another electrical device. However, in persistent cases, expert help may be required to identify the precise source and minimize interference on the electrocardiogram. Simple measures, such as proper attention to basic principles of electrocardiographic measurement, can eliminate some artifacts.

Clinical signs, along with monitored waveforms from other simultaneously monitored parameters, may provide the clues to differentiate artifacts from true changes on the electrocardiogram. 2.As can be seen from this figure d i 1 forms the primary input of the first adaptive noise canceller, d i 1 contains the ECG signal with four additive artifacts, viz. These artifacts may be nonspecific or may resemble serious arrhythmia. A proposed cascaded 4-stage adaptive noise canceller for removing the four artifacts from the ECG signal is shown in Fig. Newer diagnostic and therapeutic modalities may generate artifactual changes. Electrical devices used in the clinical setting can induce artifacts by various different mechanisms. Artifactual signals, which corrupt the normal cardiac signal, may arise from internal or external sources. Interference of the monitored or recorded electrocardiogram is common within operating room and intensive care unit environments.
